Medical laboratory electrolyte analyzer Technical Data
Specification
Sample: serum, plasma, whole blood and dilute urine
Storage: up to 10000 test results
Printer: Internal thermal printer
Interface: RS232 port
Analysis Method: ion selective electrode (ISE)
Measuring speed: ≤30s
Sample position: 38 positions (including 5 emergency and 3 functional positions)
Medical laboratory electrolyte analyzer Working environment
Temperature: 5-40°C
Relative Humidity: ≤80 %
Atmospheric pressure: 86~106 kPa
Power supply: AC220V±22V,50HZ±1HZ
Power: ≤120W
Dimension: 625mm (L)*280mm(w) *384mm(H)
Net weight: 6Kg
Medical laboratory electrolyte analyzer Measure Data
Items Measuring Range Resolution Measuring Precision(CV%)
K+ 0.50-20.00mmol/L 0.01mmol/L ≤1.5%
Na+ 15.0-200.0mmol/L 0.1mmol/L ≤1.5%
Cl- 15.0-200.0mmol/L 0.1mmol/L ≤1.5%
Ca++ 0.10-6.00mmol/L 0.01mmol/L ≤2.0%
pH 4.00-9.00pH 0.01pH ≤2.0%
Li+ 0.10-6.00mmol/ 0.01mmol/L ≤3.0%
Mg++ 0.10-4.00mmol/L 0.01mmol/L ≤3.5%
TCO2 2.0-70.0mmol/L 0.1mmol/L ≤3.5%
Medical laboratory electrolyte analyzer Main Features
Human-machine interactive menu; Dynamic and real-time display of sample ID.
Liquid level automatic detection and alarming. Real-time diagnostic of system working status.
Automatically detect and filter tiny bubbles to avoid dog and ensure accurate measurement.
Wave theory flushing method and direct flushing pipe method to avoid block and crossed contamination.
Automatic calibration and two-point correction to adjust slope and intercept; Available to print-out QC graph and QC statistical parameters.
Power failure protection to avoid data losing; Supporting fuzzy query.
Supporting LIS software; Upload data format is selectable; Supporting RTC clock management
Available to power off any time, therefore reduce reagent consumption, suitable for any hospital.